Manufacturer: FTS Systems
Model: XE753
Condition: For Parts or Not Working. This system is in Fair cosmetic condition. There are lots of small scuffs/scratches/dings especially in the bigger cabinet that make it look a little rough from close up. The data label for the base unit is missing, the shown big label (in the last picture) is for a different unit to give you an idea of what this one would look like, if it had a label on the base unit. The photos show the exact item(s) the buyer will receive.
Notes: This unit is NOT working properly. We suspect there may be water in the cooling system, which may be able to be fixed just by running VERY dry air through (-80ºC freezing rated) for a while, we are not sure. The cooling does not reach maximum cold temperature, it only gets to about -65 or so degrees. The heater overshoots badly and can only attain about 200 degrees. Eventually the airflow gets less and less as (we suspect) the water inside freezes up and chokes off the passage of air, even to the point of stopping airflow output altogether. The status lights on the controller don't light up. No other testing was done. No manuals, cords/cables, or additional items are included if not listed or shown.
Included:
- (1) Refrigeration Unit with Output Hose attached
- (1) Controller
- (1) Connection Cable
Specifications:
- Manufacturer: FTS Systems
- Model: XE753_01
- Air-Jet XE Precision Temperature Cycling System
- Temperature range: -75ºC to +225ºC
- Thermal forcing unit fits on the benchtop
- Self-contained only requires dry air, NO nitrogen required!
- Monitor and control from an internal or external (sample-mounted) thermocouple
- Save multi-step profiles in the Watlow F4 digital controller
- RS-232 or GPIB (IEEE_488) programmability to interface with automated test equipment
- Flow rates ajustable from 1.0 to 4.0 SCFM (2.0 SCFM gives best temperature control range)
- Small enough to be rack-mounted
- Refrigeration unit dimensions: 24" x 15" x 16"H (approximately)
- Built-in overtemperature safety
- Built-in low-flow safety
- PID digital controller
- Rated for +/-0.5º stability and repeatability (please read testing notes!)
- Control unit dimensions: 10" x 8" x 7" (approximately)
-
Transition specifications @ 2SCFM (please read testing notes!)
- -55ºC to +125ºC <10 seconds
- +125ºC to -55ºC <20 seconds
- Power: 220V 50Hz (compatible with 230V 60Hz)
- Requires clean dry air, 7SCFM with -80ºC dewpoint
- NO thermal caps (sample enclosures) are included
AS-IS
- No returns - No refunds - No exchanges
- By purchasing this item, you are agreeing to not seek reimbursement should the item not function to the manufacturer's specifications.